Multicenter StudyMulticenter study of oxygen-insensitive handheld glucose point-of-care testing in critical care/hospital/ambulatory patients in the United States and Canada.
Existing handheld glucose meters are glucose oxidase (GO)-based. Oxygen side reactions can introduce oxygen dependency, increase potential error, and limit clinical use. Our primary objectives were to: a) introduce a new glucose dehydrogenase (GD)-based electrochemical biosensor for point-of-care testing; b) determine the oxygen-sensitivity of GO- and GD-based electrochemical biosensor test strips; and c) evaluate the clinical performance of the new GD-based glucose meter system in critical care/hospital/ambulatory patients. ⋯ The performance of GD-based, oxygen-insensitive, handheld glucose testing was technically suitable for arterial specimens in critical care patients, cord blood and heelstick specimens in neonates, and capillary and venous specimens in other patients. Multicenter findings benchmark the performance of bedside glucose testing devices. With the new +/-15 mg/dL --> 100 mg/dL --> +/-15% accuracy criterion, point-of-care systems for handheld glucose testing should score 95% (or better), as compared with the recommended reference method. Physiologic changes, preanalytical factors, confounding variables, and treatment goals must be taken into consideration when interpreting glucose results, especially in critically ill patients, for whom arterial blood glucose measurements will reflect systemic glucose levels.