The MR image artifacts caused by minute metallic particles were investigated by imaging small powdered iron quantities from 0.01 mg to 1.7 mg in water phantoms. Images with T1-weighted GRE 3-D and T2-weighted SE 2-D sequences were reconstructed with 5 MR imagers: at 0.04 T, 0.1 T (2 scanners), 1.0 T and 1.5 T. ⋯ It appeared to behave independently on the MR imager system for all iron samples. This study shows that even microscopic magnetic particles cause a notable distortion in the MR image independently of the MR equipment used.