Hip international : the journal of clinical and experimental research on hip pathology and therapy
-
Resurfacing arthroplasty has fallen out of favour in recent years due to unfavourable survivorship in joint registries and alarming reports of soft tissue reactions around metal on metal prostheses. Our aim was to assess the effect of head size, implant design and component positioning on metal production by resurfacing arthroplasties. We measured whole blood cobalt and chromium and component position in matched populations implanted with two designs of resurfacing arthroplasty over a two-year period. ⋯ Small diameter metal resurfacing components result in increased metal generation compared with larger components. As increased metal generation has been correlated to wear and therefore failure, caution must be used on implantation of smaller components and indeed, in those who require smaller components, alternative bearing materials should be considered. These results contrast with recent findings which have demonstrated early failure for larger diameter stemmed metal-on-metal prostheses.