Malfunction of electronic medical support apparatus utilized in the ICU usually causes system failure. We report several occurrences of a potentially dangerous interaction between a continuous veno-venous hemodialysis (CVVHD) system and an intra-aortic balloon pump (IABP) counterpulsation device in four patients requiring both systems. The patients had acute renal failure in the face of multi-organ failure and were dependent upon the balloon pump for pressure support. ⋯ Electrocardiogram (ECG) tracings identified the abnormal stimulus and systematic review identified as potential sources for the creation of this interference static electricity buildup, piezoelectric properties of the polyvinyl chloride tubing, and, possibly but less likely, radiofrequency interference. A newer generation ECG cable and advanced cardiac rhythm recognition software (CardioSync) have been introduced with the Datascope System 98, and the ECG interference, although still occasionally present, does not cause erratic inflation and deflation of the intra-aortic balloon pump. Interference between different electrical support systems may occur, and we suggest that the systems be tested for compatibility before combined use and that older equipment be more rigorously tested for potential clinically significant interference.
-
Impairment of cerebral oxygenation in neonates and infants after hypothermic nonpulsatile cardiopulmonary bypass (CPB) support is well documented. The objectives of this study were: 1) using a neonatal piglet model to continuously monitor the regional cerebral oxygen saturation (rSO2) by near-infrared spectroscopy during pulsatile hypothermic CPB; and 2) to quantify the pulsatile flow in terms of energy equivalent pressure (EEP). After initiation of CPB, all piglets (n = 5) were subjected to 15 minutes of core cooling, reducing the rectal temperature to 25 degrees C, followed by 60 minutes of hypothermic CPB, then 10 minutes of cold reperfusion, and 30 minutes of rewarming. ⋯ This extra pressure may help to provide better regional cerebral oxygen saturation. During pulsatile CPB, there was no rSO2 deficiency in this high flow model. Near-infrared spectroscopy responded well to changes in rSO2 during different stages of these experiments and might be a helpful tool for intraoperative monitoring.