Journal of shoulder and elbow surgery
-
J Shoulder Elbow Surg · Sep 2001
The effect of articular malposition after total shoulder arthroplasty on glenohumeral translations, range of motion, and subacromial impingement.
The articular surface of the normal humeral head has a variable posterior and medial offset with respect to the central axis of the humeral shaft. Recreation of the normal humeral head shaft offset is postulated to be an important consideration during shoulder arthroplasty. However, the effect of humeral head malposition is unknown. ⋯ Humeral articular malposition of 4 mm or less during prosthetic arthroplasty of the glenohumeral joint may lead to small alterations in humeral translations and range of motion. Inferior malposition of greater than 4 mm can lead to increased subacromial contact; offset of 8 mm in any direction results in significant decreases in passive range of motion. Therefore if subacromial contact is to be minimized and glenohumeral motion maximized after shoulder replacement, anatomic reconstruction of the humeral head-humeral shaft offset to within 4 mm is desirable.