Bandwidths of respiratory gas flow and pressure waveforms in mechanically ventilated infants.
The frequency content of airway pressure and gas flow in mechanically ventilated infants (MVIS) has not been adequately investigated. Pressure-cycled infant ventilators generate pressure pulses with short rise-times. Gas flow is approximately equal to the derivative of pressure when lung compliance is low, and hence contains high-frequency components. ⋯ Flow bandwidths B correlated with the respiratory time constant tau (B = -77.2 tau + 26.8, R2 = 0.55, P < 0.0002), and with elastance E (B = 61.4E + 10.1, R2 = 0.74, P < 0.0001). We conclude that the bandwidth of the flow waveform increases with decreasing compliance and mechanical time constant. The frequency response of pressure and flow measurement systems should be flat +/- 10% at least up to 6 and 32 Hz respectively to obtain data with dynamic errors less than 3% in infants with low-compliance lung disease.