Comparative StudyExperimental evaluation of two new sensors for respiratory rate monitoring.
Visual observation was chosen as the reference method for measuring the respiratory rate in ten healthy volunteers. The new fibre-optic and acoustic sensors were simultaneously compared with capnography and transthoracic impedance plethysmography during normoventilation in the respiratory rate range of 6-24 breaths per minute and at a fixed respiratory rate of 13 breaths per minute. In addition a simulation of central apnoea was performed. ⋯ There was no discrepancy in the results of these methods. Each of the methods responded rapidly to an apnoeic event. The new fibre-optic and acoustic sensors correlate well with more traditional methods such as capnography and transthoracic impedance plethysmography for respiratory rate monitoring.