Physiological measurement
-
Physiological measurement · Dec 2004
Comparative StudyWhat is the adequate sampling interval of the ECG signal for heart rate variability analysis in the time domain?
Heart rate variability (HRV) analysis is considered a simple method for investigating neurocardiac regulation. However, measures of HRV may be corrupted by technical artifacts. In order to investigate the consequences of digitization errors on the time domain parameters, HRV measures from model tachograms resampled at different rates were compared. ⋯ The pNN50 shows poor accuracy and precision. An ECG sampling interval of 1 ms is recommended for HRV analysis without interpolation in order to get accurate time domain measures even in seriously reduced-variability samples. However, a lower sampling rate may be satisfactory in cases where higher variability is expected.