Physiological measurement
-
Physiological measurement · Dec 2005
The influence of cardiac preload and positive end-expiratory pressure on the pre-ejection period.
The pre-ejection period (PEP) has recently been described as a potential parameter for monitoring cardiac preload. This study further investigated the influence of changes in intravascular volume status and the application of positive end-expiratory pressure (PEEP) on the pre-ejection period. In ten pigs, ECG, arterial pressure and stroke volume derived from an aortic flowprobe were registered. ⋯ However, PEP did not change significantly (from 73 to 82 ms, P > 0.05). No correlation between the changes in PEP and changes in any other variable was observed. It is concluded that PEP is not sensitive to the changes in intravascular volume status.
-
Physiological measurement · Dec 2005
Comparative Study Clinical TrialHeart rate variability: comparison among devices with different temporal resolutions.
Several devices based on ECG can acquire beat-to-beat intervals, and some of these devices may be used for heart rate variability (HRV) analysis. Most of these devices and their methods to acquire the beat-to-beat intervals need to be validated for HRV analysis by comparing them against traditional methods. Some of these methods have low temporal resolution, which may be sufficient for certain studies. ⋯ The HRV measures were further analyzed using the one-way repeated-measures analysis of variance (ANOVA) and statistically significant differences were observed in some of the HRV parameters obtained from the low-resolution method, especially the mean RR (p < 0.001) and the mean heart rate (p < 0.001). Since the same software was used for HRV analysis of the three methods, any differences were due to the temporal resolution of the RR intervals measured by the methods. Overall, both the methods correlated very well with the referent method among healthy volunteers in resting conditions and may be used by researchers for HRV studies.
-
Physiological measurement · Dec 2005
Controlled Clinical TrialNoninvasive measurement of blood pressure variability: accuracy of the Finometer monitor and comparison with the Finapres device.
To assess the accuracy of spectral indices of arterial pressure variability and baroreflex sensitivity obtained from beat-by-beat noninvasive blood pressure recordings by the Finometer device, we compared these measures with those obtained from intra-arterial recordings. The performance of the Finometer was also compared to the traditional Finapres device. In 19 cardiac disease patients, including myocardial infarction, heart failure and cardiac transplant, we estimated the power of systolic and diastolic pressures in the VLF (0.01-0.04 Hz), LF (0.04-0.15 Hz) and HF (0.15-0.45 Hz) bands and computed absolute and percentage errors relative to intra-arterial brachial pressure. ⋯ The results for diastolic pressure were similar, but the Finapres errors in the VLF and LF bands were lower. Baroreflex sensitivity was significantly underestimated by both devices (Finometer: -31%, Finapres: -24%). Despite previous studies having shown that brachial artery waveform reconstruction performed by the Finometer has improved the accuracy of blood pressure measurement compared to the Finapres device, measurement of blood pressure variability in cardiac disease patients provides worse results in most spectral parameters and a better accuracy only in the HF band of systolic pressure.
-
Physiological measurement · Dec 2005
Clinical TrialAnalysis of intracranial pressure during and after the infusion test in patients with communicating hydrocephalus.
The cerebrospinal fluid (CSF) infusion test is used to evaluate the dynamics of CSF circulation in patients with communicating hydrocephalus and is based on constant-rate infusion of the normal saline into cerebrospinal fluid space. The aim of the study was to refine methods of the analysis of intracranial pressure (ICP) recorded during and after the infusion test. The mathematical model of cerebrospinal fluid circulation was extended by the equation describing ICP decrease after the infusion. ⋯ In 20 tests the differences between post- and pre-infusion resting ICP (DeltaICP) was higher than 1 mmHg, which was considered as significant. The mean value of DeltaICP for 20 infusion tests was 3.0 +/- 0.7 mmHg. The cerebral elasticity evaluated during the infusion was greater than the elasticity estimated from the decreasing phase after the infusion (0.24 +/- 0.07 ml(-1) versus 0.14 +/- 0.03 ml(-1); p < 0.01).
-
Physiological measurement · Dec 2005
Subtraction of 16.67 Hz railroad net interference from the electrocardiogram: application for automatic external defibrillators.
The widespread application of automatic external defibrillators (AEDs) for treating out-of-hospital cardiac arrest incidents and their particular use at railway stations defines the task for 16.67 Hz power line interference elimination from the electrocardiogram (ECG). Although this problem exists only in five European countries, it has to be solved in all AEDs, which must comply with the European standard for medical equipment requirements for interchangeability and compatibility between rail systems. The elimination of the railroad interference requires a specific approach, since its frequency band overlaps with a significant part of the frequencies in the QRS spectra. ⋯ The computation procedure is based on simple signal processing tools, which include: (i) bi-directional band-pass filtering (13-23 Hz) of the analyzed ECG segment; (ii) estimation of adequate linearity thresholds; (iii) frequency measurement and calculation of dynamic interference buffer in linear segments and (iv) phase synchronization and subtraction technique in nonlinear segments. The developed method has proved advantageous in providing sufficient quality of the output interference free ECG signal for seven defined arrhythmia types (normal sinus rhythm, normal rhythm, supraventricular tachicardia, slow/rapid ventricular tachycardia, and coarse/fine ventricular fibrillation), and simulated interferences with constant or variable frequencies and amplitudes, which cover the entire amplitude range of the input channel. The procedure is suitable to be embedded in AEDs as a preprocessing branch, which proves reliable for analysis of ECG signals, thus guaranteeing the specified accuracy of the AED automatic rhythm analysis algorithms.