Esophageal temperature monitoring during radiofrequency catheter ablation: experimental study based on an agar phantom model.
Although previous studies have established the feasibility of monitoring esophageal temperature during radiofrequency cardiac ablation using an esophageal temperature probe (ETP), some questions remain regarding its efficacy. The aims of this study were to study the effect of the location of the ETP on the temperature reached, and to test the characteristics of ETP as used in clinical practice. We constructed an agar phantom to model the thermal and electrical characteristics of the biological tissues (left atrium, esophagus and connective tissue). ⋯ In conclusion, the temperature measured by the ETP always underestimates the temperature reached in the thermocouple. This fact can be explained by the distance gap between the thermocouple and probe and by the dynamic response of the ETP. The longer the distance between the ETP and catheter axis, the higher is the temperature difference.