Physiological measurement
-
Physiological measurement · Feb 2008
Unconstrained monitoring of long-term heart and breath rates during sleep.
An unconstrained method for the long-term monitoring of heart and breath rates during sleep is proposed. The system includes a sensor unit and a web-based network module. The sensor unit is set beneath a pillow to pick up the pressure variations from the head induced by inhalation/exhalation movements and heart pulsation during sleep. ⋯ A total of 89.2% of sleep hours had a UP of not more than 25%. The profile of the heart rate revealed a periodic property that corresponded to the female monthly menstrual cycle. Our system shows promise as a long-term unconstrained monitor for heart and breath rates, and for other physiological parameters related to the quality of sleep and the regularity of the menstrual cycle.
-
Physiological measurement · Feb 2008
Image monitoring for an intraperitoneal bleeding model of pigs using electrical impedance tomography.
Current medical imaging techniques are not effective for timely detection of internal hemorrhage when the bleeding is slow and in small quantities. In this study, electrical impedance tomography (EIT) was applied to monitor the intraperitoneal bleeding of an animal model. Five healthy pigs three months old were used. ⋯ The monitoring was performed with EIT at a rate of one frame per second and continued for at least 4 h. Intraperitoneal blood volume changes could be identified by inspection of consecutive EIT images during the progression of blood injection. 30 ml of blood in the peritoneum could be detected. EIT was shown to be a promising technique for continuous monitoring of intraperitoneal bleeding over periods of time.