ECG spectral and morphological parameters reviewed and updated to detect adult and paediatric life-threatening arrhythmia.
Since the International Liaison Committee on Resuscitation approved the use of automated external defibrillators (AEDs) in children, efforts have been made to adapt AED algorithms designed for adult patients to detect paediatric ventricular arrhythmias accurately. In this study, we assess the performance of two spectral (A(2) and VFleak) and two morphological parameters (TCI and CM) for the detection of lethal ventricular arrhythmias using an American Heart Association (AHA) compliant database that includes adult and paediatric arrhythmias. Our objective was to evaluate how those parameters can be optimally adjusted to discriminate shockable from nonshockable rhythms in adult and paediatric patients. ⋯ Their performance clearly degraded with paediatric rhythms. Using a single threshold, the sensitivities and specificities were below the AHA goals, particularly VT sensitivity (60.4% for TCI and 65.8% for CM) and the specificity for other nonshockable rhythms (51.7% for TCI and 34.5% for CM). The specificities, particularly for the adult case, improve when the thresholds are independently adjusted for each adult and paediatric database.