Influence of crystalloid and colloid fluid infusion and blood withdrawal on pulmonary bioimpedance in an animal model of mechanical ventilation.
Electrical impedance tomography (EIT) is considered useful for monitoring regional ventilation and aeration in intensive-care patients during mechanical ventilation. Changes in their body fluid state modify the electrical properties of lung tissue and may interfere with the EIT measurements of lung aeration. The aim of our study was to assess the effects of crystalloid and colloid infusion and blood withdrawal on bioimpedance determined by EIT in a chest cross-section. ⋯ Colloid infusion led to a significant fall in the global end-expiratory RIC (mean differences: fluid: -91.2, p < 0.001, no-fluid: -38.9, p < 0.001), which was partially reversed after blood withdrawal (mean differences, fluid: +45.1, p = 0.047 and no-fluid: +26.2, p = 0.009). The RIC was significantly lower in the animals with additional crystalloids (mean group difference: 45.5, p < 0.001). Global and regional tidal volumes were not significantly affected by the fluid and volume states.