Signal quality indices and data fusion for determining clinical acceptability of electrocardiograms.
A completely automated algorithm to detect poor-quality electrocardiograms (ECGs) is described. The algorithm is based on both novel and previously published signal quality metrics, originally designed for intensive care monitoring. The algorithms have been adapted for use on short (5-10 s) single- and multi-lead ECGs. ⋯ A classification accuracy of 98% on the training data and 97% on the test data were achieved. Upon inspection, incorrectly classified data were found to be borderline cases which could be classified either way. If these cases were more consistently labelled, we expect our approach to achieve an accuracy closer to 100%.