Physiological measurement
-
Physiological measurement · May 2014
Forehead reflectance photoplethysmography to monitor heart rate: preliminary results from neonatal patients.
Around 5%-10% of newborn babies require some form of resuscitation at birth and heart rate (HR) is the best guide of efficacy. We report the development and first trial of a device that continuously monitors neonatal HR, with a view to deployment in the delivery room to guide newborn resuscitation. The device uses forehead reflectance photoplethysmography (PPG) with modulated light and lock-in detection. ⋯ In babies ⩾32 weeks gestation, the median reliability was 97.7% at ±10 bpm and the limits of agreement (LOA) between PPG and ECG were +8.39 bpm and -8.39 bpm. In babies <32 weeks gestation, the median reliability was 94.8% at ±10 bpm and the LOA were +11.53 bpm and -12.01 bpm. Clinical evaluation during newborn deliveries is now underway.