Comparative StudyComputerized wheeze detection in young infants: comparison of signals from tracheal and chest wall sensors.
Computerized wheeze detection is an established method for objective assessment of respiratory sounds. In infants, this method has been used to detect subclinical airway obstruction and to monitor treatment effects. The optimal location for the acoustic sensors, however, is unknown. ⋯ Comparison of wheeze rates measured over the trachea and the chest wall indicated strong correlation (r ⩾ 0.93, p < 0.001), with a bias of 1% or less and limits of agreement of within 3% for the inspiratory wheeze rate and within 6% for the expiratory wheeze rate. However, sounds from the chest wall were more often affected by disturbances than sounds from the trachea (23% versus 6%, p < 0.001). The study suggests that in young infants, a better quality of lung sound recordings can be obtained with the tracheal sensor.