Medical engineering & physics
-
Comparative Study
Monitoring of heart and respiratory rates by photoplethysmography using a digital filtering technique.
An apparatus for simultaneously monitoring heart and respiratory rates was developed using photoplethysmography (PPG) and digital filters, and compared with conventional methods. The PPG signal, which includes both heart and respiratory components, was measured at the earlobe with an original transmission mode photoplethysmographic device. A digital filtering technique was used to distinguish heart and respiratory signals from the PPG signal. ⋯ To obtain higher heart and respiratory rates, exercise was performed on an electrical bicycle ergometer. Heart and respiratory rates calculated by the new method compare to those obtained from ECG and the transthoracic impedance plethysmogram. The maximum error of heart and respiratory rates was 10 beats/min and 7 breaths/min, respectively.