A systematic review of invasive, high-fidelity pressure studies documenting the amplification of blood pressure from the aorta to the brachial and radial arteries.
It is commonly accepted that systolic blood pressure (SBP) is significantly higher in the brachial/radial artery than in the aorta while mean (MBP) and diastolic (DBP) pressures remain unchanged. This may have implications for outcome studies and for non-invasive devices calibration. We performed a systematic review of invasive high-fidelity pressure studies documenting BP in the aorta and brachial/radial artery. ⋯ Further studies on SPAmp phenotypes (positive, null, negative) are advocated. Non-invasive device calibration assumptions were confirmed, namely unchanged MBP and DBP from the aorta to the brachial artery. Data did not allow for firm conclusions on the amount of BP changes from the aorta to the radial artery, and from the aorta to the brachial/radial arteries in other populations.