The presence of implanted electronic devices with conducting leads and electrodes are contraindicated for magnetic resonance imaging (MRI), denying many patients its potential benefits. The prime concern is MRI's radio frequency (RF) fields, which can cause elevated local specific absorption rates (SARs) and potential heat injury. The purpose of this article is to develop and compare a range of passive implantable "MRI-safe" lead designs. ⋯ Lead insulation and length strongly affect implanted lead safety to RF exposure during MRI. Lead designs employing impedance and reversed winding sections offer hope for the development of passive, MRI-safe, implantable conducting leads for future human use.
-
The purpose of this article was to determine the suitability of the prostate and seminal vesicle volumes as factors to consider patients for treatment with image-guided 3D-conformal radiation therapy (3D-CRT) or intensity-modulated radiation therapy (IMRT), using common dosimetry parameters as comparison tools. ⋯ Volumes of prostate and seminal vesicles provide a reproducible and consistent basis for considering patients for treatment with image-guided 3D or IMRT plans. Patients with prostate and prostate+SV volumes <65 and 85 cm3, respectively, would be suitable for 3D-CRT. Patients with prostate and prostate+SV volumes >65 and 85 cm3, respectively, might get benefit from IMRT.