Orthop Traumatol Sur
-
Orthop Traumatol Sur · Feb 2021
No radiological and biological sign of trunnionosis with Large Diameter Head Ceramic Bearing Total Hip Arthroplasty after 5 years.
Trunnionosis of large diameter (LDH) metal-on-metal total hip arthroplasty (THA) was linked to high systemic chromium (Cr) and cobalt (Co) ion levels and local adverse reactions to metal debris (ARMD). The safety of CoC LDH THA is not yet available at mid-term. Measuring whole blood Ti level of ceramic on ceramic (CoC) LDH THA with a titanium (Ti) stem is an indirect way to assess the performance of its head-neck taper modular junction. Therefore, we wanted to determine: (1) if the whole blood Ti ion levels in patients with LDH CoC THA after a minimum of 5 years of implantation is within the expected values for similar well performing Ti THA, (2) if Ti level scientifically increases over time, which would suggest the presence of a progressive modular head/neck junction wear process, (3) if clinical or radiographical manifestations of implant dysfunction are present? ⋯ IV, retrospective study.