Ieee T Inf Technol B
-
Ieee T Inf Technol B · Mar 2003
Closed-loop glucose control in critically ill patients using continuous glucose monitoring system (CGMS) in real time.
A study was conducted to determine if continuous subcutaneous glucose monitoring (from MiniMed CGMS) could be used in real-time to control blood sugar level (BSL) in patients with critical illness. A closed-loop control system was constructed to use CGMS in a real-time manner, coupled with a proportional integral (PI) control algorithm based on a sliding scale approach, for automatic intravenous infusion of insulin to patients. A total of five subjects with high BSL (> 10 mmol/L) participated in formal studies of the closed-loop control system. ⋯ Test on equality of mean and variance for BSL attained prior to, during, and post trial showed that the controller's performance was comparable to manual control. We conclude that the automatic sliding scale approach of closed-loop BSL control is feasible in patients in intensive care. More work is needed in the refinement of the algorithm and the improvement of real-time sensor accuracy.