Nihon rinsho. Japanese journal of clinical medicine
-
Review
[A non-invasive portable blood-glucose monitoring system: sampling of suction effusion fluid].
We developed a new portable transcutaneous blood glucose monitoring system using non-invasive collection of suction effusion fluid (SEF) from human skin. A ion sensitive field effect transistor (ISFET) sensor was employed to measure glucose concentration in a very small quantity of the SEF. The system was composed of a couple of portions. ⋯ The main frame contained a dilution solution reservoir, a liquid waste reservoir, a fluid pump, a vacuum pump, a micro processor, batteries, and a user interface. The system is self-contained for portable usage during up to 6 hrs monitoring. This system may be the first blood glucose monitoring equipment which does not use blood sampling.