Objective: To explore the diagnostic value of fractionated exhaled nitric oxide (FeNO) measurement in patients with asthma-chronic obstructive pulmonary disease(COPD) overlap syndrome (ACOS). Methods: Eighty-one patients with ACOS, 76 patients with asthma, 82 patients with COPD and 39 healthy non-smoking subjects were recruited in the study. Naku Lun breath analyzer was used to measure the level of FeNO in the 4 groups. ⋯ There was no correlation between FeNO and FEV(1)% or FEV(1)/FVC in ACOS, COPD and asthma groups (r=0.12, 0.11, P>0.05; r=0.11, 0.03, P>0.05; r=0.06, 0.08, P>0.05). Conclusion: FeNO is a good marker to help clinicians differentiate ACOS from COPD. FeNO>29×10(9) was the best cutoff point for the identification of patients with ACOS from COPD.
-
Zhonghua Jie He He Hu Xi Za Zhi · Feb 2017
Comparative Study[A comparison of leak compensation in six acute care ventilators during non-invasive ventilation].
Objective: To compare the ability of leak compensation in 6 medical ventilators during non-invasive ventilation. Methods: Six medical ventilators were selected, including 3 non-invasive ventilators (V60, Flexo and Stellar150), and 3 invasive ventilators(Avea, Servo I and BellaVist). Using a lung simulator, the ability of leak compensation was evaluated during triggering and cycling in 2 respiratory mechanics conditions (high airway resistance condition and high elastance resistance condition), and each condition was performed under 2 PEEP levels (4, and 8 cmH(2)O, 1 mmHg=0.098 kPa) at 4 air leak level conditions (L0: 2-3 L/min, L1: 8-10 L/min, L2: 22-27 L/min, L3: 35-40 L/min). ⋯ Stellar150 had the shortest trigger delay time in high airway resistance condition with PEEP 4 cmH(2)O and high leak level L3[(262.8±0.8) ms , P<0.01]. V60 had a good performance on trigger delay time in high elastance resistance condition with PEEP 4 and 8 cmH(2)O, and also was most stable in increasing leak levels. Conclusion: In high airway resistance and high elastance resistance conditions with different PEEP levels and leak levels, V60, Stellar150, Flexo and BellaVista ventilators could be synchronized, among which V60, Stellar150 and Flexo presented a good performance features in specific conditions.