North American neonatal extracorporeal membrane oxygenation (ECMO) devices: 2002 survey results.
In mid 2002, surveys of active extracorporeal membrane oxygenation (ECMO) centers in the United States and Canada were conducted via E-mail regarding neonatal equipment and personnel. Seventy-four out of 99 (75%) North American ECMO centers listed in the Extracorporeal Life Support Organization (ELSO) directory responded to the survey. Of the responding centers, 95% use roller pumps, and the remaining 5% use centrifugal pumps. ⋯ At 49% of the responding centers, perfusionists were involved with the ECMO program, registered nurses were involved at 84% of the centers, and respiratory therapists were involved at 61% of the centers, perfusion assistants were involved at one center (1%), and biomedical engineers were involved at one of the centers. When compared to a 1990 survey, a shift away from using bladder boxes and toward using air bubble detectors is apparent. But other than those two shifts, ECMO is done in much the same manner as it was done 12 years ago.