Clinical impedance measurements for deep brain stimulation (DBS) electrodes in human patients are normally in the range 500-1500 Omega. DBS devices utilize voltage-controlled stimulation; therefore, the current delivered to the tissue is inversely proportional to the impedance. The goals of this study were to evaluate the effects of various electrical properties of the tissue medium and electrode-tissue interface on the impedance and to determine the impact of clinically relevant impedance variability on the volume of tissue activated (VTA) during DBS. ⋯ Impedance is often used to identify broken leads (for values > 2000 Omega) or short circuits in the hardware (for values < 50 Omega); however, clinical impedance values also represent an important parameter in defining the spread of stimulation during DBS.
-