Effect of steel and teflon infusion catheters on subcutaneous adipose tissue blood flow and infusion counter pressure in humans.
Subcutaneous tissue is an important target for drug deposition or infusion. A local trauma may induce alterations in local microcirculation and diffusion barriers with consequences for drug bioavailability. We examined the influence of infusion catheters' wear time on local microcirculation and infusion counter pressure. ⋯ ATBF and bolus mean infusion pressure increased significantly with a wear time of 48 h in Teflon but not in steel catheters. The maximal pressure required to deliver a bolus infusion increased with wear time of a catheter. A higher maximal pressure was required to deliver a bolus infusion through a Teflon than through a steel catheter. We propose that the difference in infusion counter pressure and ATBF between Teflon and steel catheters with wear time may be explained by better biocompatibility of steel than Teflon.