• Ieee T Inf Technol B · Mar 2003

    Closed-loop glucose control in critically ill patients using continuous glucose monitoring system (CGMS) in real time.

    • Frederick Chee, Tyrone Fernando, and P Vernon van Heerden.
    • School of Electrical, Electronic, and Computer Engineering, University of Western Australia, Crawley, WA 6009, Australia. chee-hf@ee.uwa.edu.au
    • Ieee T Inf Technol B. 2003 Mar 1; 7 (1): 43-53.

    AbstractA study was conducted to determine if continuous subcutaneous glucose monitoring (from MiniMed CGMS) could be used in real-time to control blood sugar level (BSL) in patients with critical illness. A closed-loop control system was constructed to use CGMS in a real-time manner, coupled with a proportional integral (PI) control algorithm based on a sliding scale approach, for automatic intravenous infusion of insulin to patients. A total of five subjects with high BSL (> 10 mmol/L) participated in formal studies of the closed-loop control system. Subjects were recruited from critically ill patients in the intensive care unit (ICU) after informed consent was obtained. Error grid analysis showed that 64.6% of the BSL readings as determined in real time using CGMS sensor, when compared to conventional BSL measurements on blood drawn from an arterial line, was clinically accurate (i.e., < 20% deviation from glucometer value). In the five patients who underwent closed-loop control, the controller managed to control only one patient's glycaemia without any manual intervention. Manual intervention was required due to the real-time sensor reading deviating more than 20% from the glucometer value, and also as a safety mechanism. Test on equality of mean and variance for BSL attained prior to, during, and post trial showed that the controller's performance was comparable to manual control. We conclude that the automatic sliding scale approach of closed-loop BSL control is feasible in patients in intensive care. More work is needed in the refinement of the algorithm and the improvement of real-time sensor accuracy.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.