Observational Study
Validation of noninvasive hemoglobin measurement by pulse co-oximeter in newborn infants.
- C Nicholas, R George, S Sardesai, M Durand, R Ramanathan, and R Cayabyab.
- 1] Division of Neonatal Medicine, Department of Pediatrics, LAC+USC Medical Center, University of Southern California, Los Angeles, CA, USA [2] Division of Neonatal Medicine, Department of Pediatrics, Children's Hospital Los Angeles, Keck School of Medicine, University of Southern California, Los Angeles, CA, USA.
- J Perinatol. 2015 Aug 1; 35 (8): 617-20.
ObjectiveTo describe the accuracy of noninvasive hemoglobin (Hb) obtained with pulse co-oximeter (SpHb) compared with total Hb (tHb) from laboratory co-oximeter in neonates.Study DesignNeonates with birth weight (BW) <3000 g admitted to LAC+USC Medical Center neonatal intensive care unit were included. SpHb was recorded using Masimo Radical-7 and compared with tHb. A total of three data sets were obtained for each patient. Regression analysis and Bland-Altman analysis were performed.ResultSixty-one patients (mean±s.d., BW 1177±610 g and gestational age 28.7±3.9 weeks) were enrolled. The mean tHb value was 13.9±2.0 g dl(-1) and the mean SpHb was 14.0±2.0 g dl(-1). There was a moderately positive correlation between SpHb and tHb (r=0.66, P<0.001) with a bias and precision of -0.09±1.67 g dl(-1). Data from a subgroup of infants with gestational age ⩽32 weeks (52/61 patients) were analyzed, and the correlation coefficient was moderately positive (r=0.69, P<0.001) with a bias and precision of -0.23±1.60 g dl(-1).ConclusionOur results suggest that noninvasive SpHb may be considered as an adjunct to invasive tHb measurements in newborn infants <3000 g especially in preterm infants ⩽32 weeks of gestation.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.