-
Comparative Study
The impact of imposed expiratory resistance in neonatal mechanical ventilation: a laboratory evaluation.
- Robert M DiBlasi, John W Salyer, Jay C Zignego, Gregory J Redding, and C Peter Richardson.
- Respiratory Care Department, Children's Hospital and Regional Medical Center, Seattle, Washington, USA. robert.diblasi@seattlechildrens.org
- Resp Care. 2008 Nov 1;53(11):1450-60.
BackgroundSmall endotracheal tubes (ETTs) and neonatal ventilators can impact gas exchange, work of breathing, and lung-mechanics measurements in infants, by increasing the expiratory resistance (R(E)) to gas flow.MethodsWe tested two each of the Babylog 8000plus, Avea, Carestation, and Servo-i ventilators. In the first phase of the study we evaluated (1) the imposed R(E) of an ETT and ventilator system during simulated passive breathing at various tidal volume (V(T)), positive end-expiratory pressure (PEEP), and frequency settings, and (2) the intrinsic PEEP at various ventilator settings. In the second phase of this study we evaluated the imposed expiratory work of breathing (WOB) of the ETT and ventilator system at various PEEP levels during simulated spontaneous breathing using an infant lung model. Pressure and flow were measured continuously, and we calculated the imposed R(E) of the ETT and each ventilator, and the intrinsic PEEP with various PEEP, V(T), and frequency settings. We measured the imposed expiratory WOB with several PEEP levels during a simulated spontaneous breathing pattern.ResultsThe ventilator's contribution to the imposed R(E) was greater than that of the ETT with nearly all of the ventilators tested. There were significant differences in ventilator-imposed R(E) between the ventilator brands at various PEEP, V(T), and frequency settings. The Babylog 8000plus consistently had the lowest ventilator-imposed R(E) in the majority of the test conditions. There was no intrinsic PEEP (>1 cm H(2)O) in any of the test conditions with any ventilator brand. There were also no significant differences in the imposed expiratory WOB between ventilator brands during simulated spontaneous breathing.ConclusionsThe major cause of R(E) appears to be the ventilator exhalation valve. Neonatal ventilators that use a set constant flow during inhalation and exhalation appear to have less R(E) than ventilators that use a variable bias flow during exhalation. Clinical studies are needed to determine whether the imposed R(E) of these ventilator designs impacts gas exchange, lung mechanics, or ventilator weaning.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.