• Ann Fr Anesth Reanim · Jan 2001

    Case Reports

    [Intraoperative ventilatory obstruction by twisting of a reusable coiled expiratory breathing tube].

    • H Cuche, J Bientz, and J C Otteni.
    • Service d'anesthésie-réanimation, hôpital Edouard-Herriot, place d'Arsonval, 69437 Lyon, France.
    • Ann Fr Anesth Reanim. 2001 Jan 1;20(1):54-6.

    AbstractA case of intraoperative subtotal obstruction of a reusable coiled expiratory breathing tube is reported. Partial occlusion by twisting was made possible by detachment of the coil from the external face of the tube after multiple reprocessings with high drying temperatures. A technique for tube checking before reuse is described.