-
Ann Fr Anesth Reanim · Jan 2001
Case Reports[Intraoperative ventilatory obstruction by twisting of a reusable coiled expiratory breathing tube].
- H Cuche, J Bientz, and J C Otteni.
- Service d'anesthésie-réanimation, hôpital Edouard-Herriot, place d'Arsonval, 69437 Lyon, France.
- Ann Fr Anesth Reanim. 2001 Jan 1;20(1):54-6.
AbstractA case of intraoperative subtotal obstruction of a reusable coiled expiratory breathing tube is reported. Partial occlusion by twisting was made possible by detachment of the coil from the external face of the tube after multiple reprocessings with high drying temperatures. A technique for tube checking before reuse is described.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.