Comparative Study
Biomechanical testing of a novel four-rod technique for lumbo-pelvic reconstruction.
- Brian P Kelly, Francis H Shen, John S Schwab, Vincent Arlet, and Denis J Diangelo.
- Department of Biomedical Engineering and Imaging, University of Tennessee Health Science Center, Memphis, TN 38163, USA. bkelly@utmem.edu
- Spine. 2008 Jun 1;33(13):E400-6.
Study DesignA biomechanical testing protocol was used to study different lumbo-pelvic fixation techniques in a human cadaveric lumbar spine model.ObjectiveTo compare the in vitro biomechanics of a novel four-rod lumbo-pelvic reconstruction technique with and with out cross-links, to that of a conventional cross-linked two-rod technique.Summary Of Background DataNumerous lumbo-pelvic reconstruction methods based on the Galveston two-rod technique have been proposed for cases involving total sacrectomy. Recently a technique that proposes novel use of 4 supporting longitudinal rods across the lumbo-pelvic junction has been reported. No comparative in vitro biomechanical testing has been previously done to evaluate these different reconstruction methods.MethodsFive spines were evaluated in flexion, extension, left-right lateral bending and left-right axial rotation in a human total sacrectomy model. The model was comprised of cadaveric lumbar spines (L1-L5) with custom fabricated polyethylene blocks used to simulate pelvic fixation. Three conditions were evaluated: Linked Four-Rod, Linked Two-Rod, and Four-Rod (no cross-links). Flexibility and motion data were compared using a one-way repeated measures analysis of variance and SNK tests.ResultsThe Linked Four-Rod and Four-Rod conditions significantly decreased flexibility and reduced L5-Pelvic motion over the Linked Two-Rod construct in flexion and extension. The Linked Four-Rod condition significantly decreased flexibility in left-right axial rotation compared with the Four-Rod and Linked Two-Rod conditions. No significant differences occurred in relative lateral movement between left and right pelvic polyethylene blocks.ConclusionThe four-rod technique improved fixation stability over the conventional linked two-rod technique in flexion and extension, and when cross-linked, in left-right axial rotation. The four-rod technique also significantly reduced L5-Pelvic junction movement in flexionand extension, which may have implications for bony fusion. The use of cross-links is recommended.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.