Journal of biomechanical engineering
-
Currently, rigid fixation systems are the gold standard for degenerative disk disease treatment. Dynamic fixation systems have been proposed as alternatives for the treatment of a variety of spinal disorders. These systems address the main drawbacks of traditional rigid fixation systems, such as adjacent segment degeneration and instrumentation failure. ⋯ In extension, the facet load in DSDR case was similar to that in intact spine. The dynamic rod constructions (DSDR and RSDR) led to a lesser peak stress at screws compared with rigid rod constructions (DSRR and RSRR) in all loading cases. A dynamic construct consisting of a dynamic rod and a dynamic screw did protect the adjacent level from excessive motion.