Errors of oscillometric blood pressure measurement as predicted by simulation.
To study factors affecting the accuracy of oscillometric measurement. ⋯ Oscillometric readings of systolic and diastolic pressures are strongly influenced by pulse pressure and the shape of the artery-cuff pressure/volume curve, whereas those of the mean arterial pressure are affected by pulse pressure and both the shape of the artery-cuff pressure/volume curve and the shape of the arterial pulse.
-
Blood pressure monitoring · Oct 2011
Application of Nexfin noninvasive beat-to-beat arterial blood pressure monitoring in autonomic function testing.
Evaluation of autonomic function responses is increasingly important for risk prediction and hemodynamic evaluation in the ambulant and perioperative setting, but requires a noninvasive arterial blood pressure measurement device. This study describes whether a novel noninvasive beat-to-beat arterial blood pressure measurement device (Nexfin HD) is able to reproducibly reflect autonomic function responses in healthy volunteers. ⋯ This study shows that noninvasive beat-to-beat Nexfin HD arterial blood pressure measurements reproducibly reflect autonomic function responses in healthy volunteers.